FIBER Q
M
UICK
Settings
To change the settings on the troubleshooter:
Press
1
To select a setting to change, press
2
Use
3
Press
4
BACKLIGHT
The display backlight turns off if you do not press a key for the
period of time shown. You can set the time to 15 to 60 seconds
in increments of 5 seconds.
I.O.R. (index of refraction)
The index of refraction is the ratio of the speed of light in a
vacuum to the speed of light in a fiber. The troubleshooter uses
the index of refraction to calculate length. If you increase n, the
calculated length decreases. The default is 1.496. This value is
satisfactory for most fibers.
ft/m
Select feet or meters as the unit for length measurements.
LOSS LIMIT
Sets the minimum value of power loss for incidents that the
troubleshooter shows an incident. You can select a value from
0.50 dB (lower loss) to 6.10 dB (higher loss) in 0.2 dB increments.
The default is 0.70 dB.
When you change the loss limit, the troubleshooter
also applies the new limit to the results shown from a
previous test.
8
Users Manual
AP
.
to change the setting.
or
to save the setting.
Note
, then press
.
Need help?
Do you have a question about the FIBER QUICKMAP and is the answer not in the manual?