The option displayed when exiting the settings screen is saved and used for
future measurements.
The following is a description of each setting.

Gas Settings

This setting tells the Analyzer what type of gas is flowing through the high- and low-flow
ports. The Analyzer uses this information to improve the accuracy of the flow
calculations.
Failure to enter the correct value causes the flow and volume calculations
to be in error.
To modify the gas settings:
1. From the Setup menu, highlight Settings and press the ENTER soft key.
2. Highlight Gas Settings and press the MODIFY soft key.

Gas Type

The possible values for Gas Type are shown in Table 2-1.
N2O bal O2
User Defined User selects custom gas mixture. This option requires the VT for Wondows
For balance gas settings, the oxygen concentration is measured using the
Analyzer oxygen sensor located on the high-flow port. If the low-flow port
is to be used, allow gas to flow through both high- and low-flow ports so
that the oxygen can be measured.
To modify the gas type:
2-8
Depending on the option, a value can be pre-selected or input from the keypad. The
following is a description of each of the options.
Value
Air
Standard room air
N2
100 % Nitrogen
N2O
100 % Nitrous Oxide
CO2
100 % Carbon Dioxide
O2
100 % Oxygen
Heliox
30 % Oxygen and 70 % Helium blend
Measured Oxygen balance Nitrous Oxide
He bal O2
Measured Oxygen balance Oxygen Helium
N2 bal O2
Measured Oxygen balance Nitrogen
PC software.
Note
Note
Table 2-1. Gas Type Values