PDC 2000/3000 Repair Manual
7. Set test parameters as follows: (Left side of VI front panel)
• Do no tests to OFF
• Do all tests to ON
• Load defaults to ON
8. Click on the white arrow. Arrow should change to black.
9. Click on the green Start button to automatically start the selected LabView VI test
program. The EEPROM calibration data panel appears.
Note:
If the camera is not turned on, turn on the camera under test and then
repeat step 7 through 9.
10. Click on the Start button to automatically start the ambient part of this calibration
procedure.
Note: Follow all instructions that appear on the screen while the ambient part of this
calibration procedure runs. Click on the appropriate responses that appear in
the dialog boxes.
11. Upon completion of the ambient part of the calibration procedure, the IR CAL.VI panel
appears. This panel prompts you to move the camera under test to the Graywall Booth
(Figure 3-3).
29
Testing&Adjustment