(II)
M+/SAVE
ZERO/ESC
4) Press key
to confirm and save. Press key
to return to weighing
mode without saving.
Place the sample on the weighing pan.
HI indication appears, when the item on the weighing pan is greater than the upper limit
OK indication appears, when the item on the weighing pan is between upper and lower limits.
LO indication appears, when the item on the weighing pan is smaller than lower limit
Note: the item on the weighing pan should be more than or equal to 20e.
When changing the Hi-Lo value, the scale will activate the weighing checking function
automatically. If the Lo value is higher than Hi value, then the Hi value will become the
same value as Lo data.

7-4 Basic Counting

7-4-1 Entering a Known Unit Weight and Unit weight Memory

1) Under the weighing mode, use the numeric
"·"
keys together with key
to input unit weight
UWS
.
value, and then press
2)Put the items on the weighing pan , the
scale starts counting.
Note: Symbol " " points at "UW" when entered unit weight is lower than 4/5 of scale division.
Unit weight Calling-out
UWS/ITEM
1) Press
key. The window displays as shown on the right.
TARE/
/MR
M-/HI
2) Input the number of unit weight. Press key
or
to shift and key
or
PRINT/LO
to modify the value (or via numeric keys)
M+/SAVE
3) Press key
to confirm called-out weight and return to the weighing mode. Press
ZERO/ESC
key
to exit and return to weighing mode.

7-4-2 Sample Counting and ACAL

Sample Counting
1) Place samples onto the weighing pan (or into a tared container) and input the quantity
SMPL
with numeric keys, then press key
.
Symbol " " pointing at "ACAI"
The unit weight is calculated, with
.
12