Connecting Voltage/Current Input Pods

Connecting Voltage/Current Input Pods
Two Series 5500 input pod types, voltage and current, can connect alternatively to the
Pod types
same Voltage/Current Input Module Model No. 61MZP with 25-PIN interface cable
connector. The inputs are attenuated to low voltage signals that can be measured by the
61000.
This equipment has been tested and found to comply with emissions and/or immunity
Caution
requirements. The protective earth ground of the 61000 mainframe and the voltage/
current pods must be connected to the same ground reference. Failure to do so is likely
to result in undesired interference.
Photos of Model 5536 voltage pod and Model 5537 current pod are shown below. See
Pod assembly
diagram
pod assembly labels on page 2-29.
2-28
1
2
3
SF-212
3
SF-213
Continued on next page