Connecting an External DC Power Supply
Introduction; Connecting the DC Input Cable

Introduction

An external dc source of power can be connected to the 61000 for normal operation.
Introduction
This may be required where ac power is not available or operation from the internal
battery may extend past the batteries capability.
An external stand-alone +12Vdc power supply of at least 5 amps capacity may be used.
Connection is made to the input power connector through an optional dc cable
assembly.
Physical injury or damage to the instrument could result from incorrect
WARNING
connections. Always observe correct polarity. Make connections to the 61000 first
and then make connections to the external dc power source.
Podrían producirse lesiones físicas o daños al instrumento debido a conexiones
ADVERTENCIA
incorrectas. Siempre observe la polaridad correcta. Efectúe primero las
conexiones al 61000 y luego haga las conexiones a la fuente de energía de corriente
continua externa.
Des connexions incorrectes risquent de causer des blessures aux personnes ou
AVERTISSEMENT
l'endommagement du matériel. Observez toujours la polarité correcte. Faites les
connexions au 61000 en premier, puis branchez sur la source d'alimentation c.c.
externe.
Falsche Verbindungen können zu Verletzungen oder Schäden am Gerät führen.
WARNUNG
Immer auf richtige Polarität achten. Zuerst Anschlüsse mit dem 61000 und dann
Anschlüsse mit der externen Gleichspannungsquelle vornehmen.
Internal 12V dc circuitry is protected by resettable fuses mounted on the internal power
Fuses
supply. These fuses will operate when an excessive external 12V voltage is applied or
when a fault condition within the internal circuit is experienced. To reset the fuse, turn
power off to the 61000, disconnect the 12V power cord, wait 5 seconds, then reconnect.