Connecting to AC Power Source
For 61000 Standard (61STD) and Single Module (61VCM) Units
Connecting to
an AC power
source
Follow these steps to connect to an ac power source
Step
For Switchgear (61SG and 61SGD) Mounting Units
The mains supply power to the Encore Series 61SG or SGD must be installed
Safety
Disconnects
downstream from a switched current limiting device. The circuit protection device
should be 20 Amps or less, and must be rated for the available voltage and fault
current; 5 Amps fuses are preferred.
All mains supply conductors connected to the Encore Series 61SG or SGD must
WARNING
originate at circuit breakers or fuses rated 20 Amps or less.
You must provide a method for manually removing power from the Encore Series
61SG or 61SGD such as a clearly labeled circuit breaker or a fused disconnect
switch.
Step
2-8
1
Set the power switch to OFF.
2
Verify the operating voltage range for your 61000, as marked on the
power input decal.
3
Connect the appropriate power cord for the voltage rating of the unit.
Make connection to the instrument first.
4
Connect power cord to the power source.
5
Turn power switch to ON.
1
With all power OFF, connect the green ground wire to Safety (Earth)
ground first before making any other connections.
2
Connect the positive (+) terminal of the mains supply to the positive (+)
terminal of the instrument.
3
Connect the negative (-) terminal of the mains supply to the negative (-)
terminal of the instrument.
4
Turn power switch to disconnect device to ON.
5
Turn power switch of the instrument to ON position.
.
Action
Action