Battery Status Display; Conditioning The Battery - Fluke DSP-4100 User Manual

Cableanalyzer dsp-4000 series
Hide thumbs Also See for DSP-4100:
Table of Contents

Advertisement

Battery Status Display

To see the charge level of the main unit's NiMH battery, turn the rotary switch to
SPECIAL FUNCTIONS; then select Battery Status. To see the charge
level of the remote's battery, connect the remote to the main unit; then use ! to
toggle the display.

Conditioning the Battery

The battery status is more accurate when the battery is properly conditioned. The
test tool includes a battery conditioning feature. To condition the batteries, proceed
as follows:
1. Connect the main and remote units together with a calibration module, or with
LIA adapters and an RJ45 patch cord or a cable.
2. Connect ac adapters to both units.
3. Turn on the remote.
4. Press and hold D while turning on the main unit. Release the key when the
display goes blank.
The conditioning function discharges then charges the battery. This may take
several hours, depending on the battery status when conditioning starts. You can
interrupt the conditioning if necessary. The ac power LEDs on the test tool and
remote turn green when conditioning is complete.
Getting Started
Battery Status Display
2
2-45

Advertisement

Table of Contents