During the configuration process, you can abandon the changes you have made by
pressing ! Abort.
To reset a parameter back to the default value, highlight the parameter; then press
@ Set to Default.
To reset all of the parameters back to default values, highlight the test standard on
Page 1 of the configuration screens; then press E. Use U D to select the
desired default standard; then press E.
To configure a custom test, proceed as follows:
1. Turn the rotary switch to SETUP.
2. Press $ Page Down until you see the CONFIGURE CUSTOM TEST
selection.
3. Highlight a custom test name. The default cable names are * Custom
Cable 1* through * Custom Cable 8 *. An asterisk always
proceeds and follows the name of a custom test configuration. Press E.
4. Use U D to select a parameter to change. Use $ Page Down to see
additional parameters. Press E to select a highlighted parameter for
editing.
Selecting a default standard for a previously configured custom test
resets all test parameters to default values.
5. Use the appropriate keys to select the value you want; then press E.
If you selected to change the custom test name or the cable type, the
alphanumeric entry screen appears. Use the appropriate keys to enter a name
up to 27 characters long. When you are finished entering characters, press
S.
6. Repeat steps 4 and 5 to change additional parameters.
7. When you are finished configuring test parameters, press S to store the
custom configuration.
Calibrations and Custom Test Standards
Configuring a Custom Test
Note
6
6-5