Output Resistance
Line regulation
±(%of Output+Offset)
Load regulation
±(%of Output+Offset)
Setup Resolution
Read Back Resolution
Setup Accuracy
( within 12 months, 25℃
±5℃)
±(%of Output+Offset)
Read Back Accuracy
( within 12 months, 25℃
±5℃)
±(%of Output+Offset)
Ripple
(20Hz -20MHz)
Setup Temperature
Coefficient
(%of Output/℃+Offset)
Read Back Temperature
Coefficient
(%of Output/℃+Offset)
Rise time( no load )
Rise time( full
5
load )
Fall time( no load )
Fall time( full load )
Current seamless
6
switching
Transient Response
Time
7
AC Input
Setup
stability-30min
( %of Output +Offset )
Setup stability-8h
( %of Output +Offset )
Readback
stability-30min
( %of Output +Offset )
Readback stability-8h
( %of Output +Offset )
efficiency
Remote Sense
External Sink
1
Power
Range
2
Accuracy
Resolution
Voltage
Current
Voltage
Current
Voltage
Current
Power
Voltage
Current
Power
Voltage
Current
Power
Voltage
Current
Power
Voltage
Current
Voltage
Current
Voltage
Current
5
Voltage
Voltage
5
Voltage
5
Voltage
-90%-90%
90%--90%
Voltage
Voltage
wire)
Frequency
Voltage
Current
Voltage
Current
Voltage
Current
Voltage
Current
Copyright© Itech Electronics Co., Ltd.
0-63KW
0-1.904Ω
0.25%+100mΩ
≤0.01%+20mV
≤0.01%+210mA
≤0.01%+50mV
≤0.05%+420mA
100mA
100mA
≤0.05%+100mV
≤0.2%+420mA
≤1%+210W
≤0.05%+100mV
≤0.2%+420mA
≤1%+210W
≤200mVp-p
≤350mArms
≤0.05 %+100mV
≤0.2%+420mA
≤0.05%+100mV
≤0.2%+420mA
≤15ms
≤30ms
≤50ms
≤15ms
≤70ms
≤70ms
≤2ms
≤0.05%+100mV
≤0.2%+420mA
≤0.05%+100mV
≤0.2%+420mA
≤0.05%+100mV
≤0.2%+420mA
≤0.05%+100mV
≤0.2%+420mA
89%(50V/420A)-90.5%(200V/105A)
2V
Specification
1mΩ
10mV
0.1W
10mV
0.1W
380V±10%(Three-phase
47Hz-63Hz
five
94
Need help?