The menu » Settings «
Use the »
« and »
M
key to confirm the setting.
Select » Next « and press the »
— The settings of the A1 are indicated.
To confirm select » Confirm « and press the »
— The settings for alarm A1 are now complete.
– Set alarm A2
— This function is used to change the alarm configuration for alarm A2.
Select the menu items "Settings «, "Instrument «, "Alarm « and "Alarm A2 « in this
order, pressing the »
OK
— The settings are made in the same manner as for alarm A1.
– Set ack.
— This function is used to enable or disable the acknowledgement of the alarms with
the »
« key.
OK
Select the menu items » Settings «, » Instrument «, » Alarm « and » Set ack. « in
this order, pressing the »
Select » Enable « or » Disable « and press the »
– Test alarm A1
— This function simulates the A1 alarm state for testing purposes.
Select the menu items » Settings «, » Instrument «, » Alarm « and » Test alarm A1 «
in this order, pressing the »
— The function » Test A1 status « is opened.
Select » Enable « or » Disable « and press the »
— When the function is enabled, the relay and the interfaces are set to the
A1 alarm state.
Page 46 of 72 Pages
« keys to set each digit of the value and press the »
« key.
OK
« key.
OK
« key after each selection
« key after each selection.
OK
« key to activate it.
OK
« key after each selection.
OK
« key to activate it.
OK
«
OK