Seft Calibration; Self Calibration - Hantek 1008 User Manual

Digital
Hide thumbs Also See for 1008:
Table of Contents

Advertisement

3. Attach the tip of probe and ground tab to the Connector of Probe compensator. Click
the button. A square wave will be displayed in several seconds. (Approximately
1 kHz, 2V, peak- to- peak).
4. Inspect CH2-CH8 with the same method. Repeat step 2 and 3.

Self Calibration

The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5 deg or more. For accurate calibra-
tion, power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the "Utility -> Calibration" option and follow the directions on the screen.
The self calibration routine takes about several minutes.