Hantek 6022BE User Manual
Hide thumbs Also See for 6022BE:

Advertisement

USER'S MANUAL
Hantek6022BE
V1.0.5
www.hantek.com

Advertisement

Table of Contents
loading

Summary of Contents for Hantek 6022BE

  • Page 1 USER’S MANUAL Hantek6022BE V1.0.5 www.hantek.com...
  • Page 2: Table Of Contents

    Content General Safety Summary ................1 Chapter 1 Getting Start.................. 2 1.1 System Requirement ....................3 1.2 Install Software ......................4 1.3 Install Driver ....................... 7 1.4 General Features ..................... 11 1.5 General Check ......................12 1.6 Probe Compensation ....................12 1.7 Function Check ......................
  • Page 3: General Safety Summary

    Hantek 6022BE General Safety Summary Review the following safety precautions carefully before operate the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards use the device as specified by this user’s guide only.
  • Page 4: Chapter 1 Getting Start

    Hantek 6022BE Chapter 1 Getting Start The oscilloscope is small, lightweight, no external power required, portable oscilloscopes! The oscilloscopes is ideal for production test, research and design and all of the applications involving analog circuits test and troubleshooting, as well as education and training.
  • Page 5: System Requirement

    Hantek 6022BE 1.1 System Requirement To run the oscilloscope software, the needs of computer configuration are as follows: Minimum System Requirements Operating System Window XP/Vista/Win7/Win8/Win 10 Processor Upwards of 1.00G processor Memory 128M byte Disk Space 500M disk free space...
  • Page 6: Install Software

    Hantek 6022BE 1.2 Install Software Caution: You must install the software before using the oscilloscope. While in Windows, insert the installation CD into the CD-ROM drive. The installation should start up automatically. Otherwise in Windows Explorer, switch to the CD-ROM drive and run Setup.exe.
  • Page 7 Hantek 6022BE Choose a destination directory. Click 'Next' to continue. Check the setup information. Click Next to start copying of files. User’s Manual...
  • Page 8 Hantek 6022BE This Status dialog is displayed during copying of files. Updating Your System Configuration. The installation is complete. User’s Manual...
  • Page 9: Install Driver

    Hantek 6022BE 1.3 Install Driver Connect the A-Type Plug of USB cable to your PC’S USB port. Connect the B-Type Plug of USB cable to USB port. New hardware is found. New hardware search wizard starts. User’s Manual...
  • Page 10 Hantek 6022BE 5. New hardware search wizard starts to search the driver. 6. New hardware wizard installs “DRIVER”. User’s Manual...
  • Page 11 Hantek 6022BE 7. The wizard has finished installing for “DRIVER”. 8. New hardware is found. User’s Manual...
  • Page 12 Hantek 6022BE 9. Found new hardware Wizard. Select “Install the software automatically”. 10. New hardware wizard installs software and finish new hardware search wizard. User’s Manual...
  • Page 13: General Features

    Hantek 6022BE 11. Your new hardware is installed and ready to use. 1.4 General Features  Maximum real-time sample rate: 48MSa/s  Memory depth: 1Mbytle/channel  Built-in Fast Fourier Transform function(FFT);  20 Automatic measurements;  Automatic cursor tracking measurements;...
  • Page 14: General Check

    Hantek 6022BE 1.5 General Check Please check the instrument as following steps after receiving an oscilloscope:  Check the shipping container for damage: Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
  • Page 15: Function Check

    Hantek 6022BE Over compensated Under Compensated ■ If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the fattest square wave being displayed on the oscilloscope. ■ Repeat if necessary. WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
  • Page 16 Hantek 6022BE  Input a signal to a channel of the oscilloscope The oscilloscope is equipped with dual channels. Please input signal in the following steps: Set the attenuation switch on the probe as 10X and connect the probe on the oscilloscope with CH1.
  • Page 17: Self Calibration

    Hantek 6022BE 1.8 Self Calibration The self calibration routine lets you optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but you should always run the routine if the ambient temperature changes by 5℃ or more. For accurate calibration, power on the oscilloscope and wait twenty minutes to ensure it is warmed up.
  • Page 18: Chapter 2 Operating Basics

    Hantek 6022BE Chapter 2 Operating Basics  The User’s Interface  The Menu System  The Vertical System  The Horizontal System  The Trigger System  Input Connector User’s Manual...
  • Page 19: The User's Interface

    Hantek 6022BE 2.1 The User’s Interface Click the software icon on the desk after you finished the software setting and equipment connecting. Then a user interface will be showed as follows: In addition to displaying waveforms, the display area is filled with many details about the waveform and the oscilloscope control settings.
  • Page 20 Hantek 6022BE 10. Displays the CH2 information Readouts show the vertical scale factors of the channels. A “B” icon indicates that the channel is bandwidth limited. 11. Displays the CH1 information Readouts show the vertical scale factors of the channels.
  • Page 21: The Menu System

    Hantek 6022BE 2.2 The Menu System The Main Menu File: Load or Save data, setup View: Change the user interface Horizontal: Set the horizontal system Vertical: Set the vertical system User’s Manual...
  • Page 22 Hantek 6022BE Trigger: Set the trigger system Display: The waveform display setup Cursor: Cursor measure setup Utility: Utility setting Help: Turn on help file User’s Manual...
  • Page 23: The Vertical System

    Hantek 6022BE 2.3 The Vertical System Click “Vertical” in main menu The following figure shows the vertical system menu. It shows the vertical parameters setting. CH1: Leads to the setting of the CH1 parameters CH2: Leads to the setting of the CH2 parameters Use the left mouse button to dragging the channel icon to set the vertical position offset precisely.
  • Page 24: The Trigger System

    Hantek 6022BE 2.5 The Trigger System Click “Trigger” in main menu. Mode: Sets the trigger mode to Edge trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE Source: Selects the trigger source to CH1, CH2. Slope: Selects the edge trigger slope to Positive or Negative slope The user can drag the trigger icon to adjust the trigger level and the trigger level will be displayed on the top right corner.
  • Page 25: Chapter 3 Understanding Function

    Hantek 6022BE Chapter 3 Understanding Function  Set Oscilloscope  Set Vertical System  Set Horizontal System  Set Trigger System  Measure Signal  Utility Function  The Display System  Save/Load  Print User’s Manual...
  • Page 26: Set Oscilloscope

    Hantek 6022BE 3.1 Set Oscilloscope Save Setup The oscilloscope software saves the current setup before you close the oscilloscope software. The oscilloscope recalls this setup the next time you run the software. You can use the “Save Setup” menu to permanently save up to several different setups.
  • Page 27 Hantek 6022BE You can left click and drag the mouse on the green knob or the yellow knob to change the voltage.  Probe Attenuation Setting Select the attenuation factor for the probe. To check the probe attenuation setting, toggle the probe menu to match the attenuation factor of the probe.
  • Page 28 Hantek 6022BE Turn off invert Turn on invert The following picture shows the waveform before inversion: The following picture shows the waveform of inversion: User’s Manual...
  • Page 29 Hantek 6022BE 3.2.1 Set Math Click in Toolbar. The MATH Menu:  Turn On/Off: Turn On/Off the MATH Channel.  Source A/B: Set the sources of the math channel.  Operate: Set operates type of the math channel. The mathematic functions include addition, subtract, multiply and division for CH1 and CH2.
  • Page 30 Hantek 6022BE The Math Function Display: Fast Fourier Transform Function To use the FFT mode, you need to click in Toolbar to open the FFT setup window. The FFT setup window: User’s Manual...
  • Page 31 Hantek 6022BE You can select the Source channel, Window algorithm, FFT number, and FFT Zoom factor. It displays only one FFT spectrum at a time. Source: Selects the channel used as the FFT source Window: Selects the FFT window type...
  • Page 32: Set Horizontal System

    Hantek 6022BE You can change the vertical scale of a waveform. The waveform display will contract or expand relative to the reference level. The Reference Waveform Display Window : Note: If you turn on the “Reference” channel, the load file window will appear.
  • Page 33: Set Trigger System

    Hantek 6022BE Format: leads the setting of the horizontal format parameters The time base ranges of the oscilloscope is listed as follows. The horizontal scan speed is from 4ns/div to 1hour/div. In the “Format” menu of the “Horizontal” menu, select the display mode to Y-T, X-Y.
  • Page 34: Measure Signal

    Hantek 6022BE Auto: Acquire waveform even no trigger occurred Normal: Acquire waveform when trigger occurred. Single: Acquire waveform when trigger occurred then stop Source: You can use the trigger source options to select the signal that the oscilloscope uses as a trigger.
  • Page 35 Hantek 6022BE Type There are four types of cursors: Cross, Vertical and Horizontal. Cross The Cross cursors appear as cross lines on the display and measure the vertical and horizontal parameters. The Cross cursor display window: Vertical The Vertical cursors appear as vertical lines on the display and measure the vertical parameters.
  • Page 36 Hantek 6022BE The Vertical cursor display window: Horizontal The Horizontal cursors appear as horizontal lines on the display and measure the horizontal parameters. The Horizontal cursor display window: User’s Manual...
  • Page 37 Hantek 6022BE 3.5.2 Measure Click “Measure” in main menu. The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time measurements). Vertical Maximum: Voltage of the absolute maximum level, Measured over the entire waveform Minimum: Voltage of the absolute minimum level, Measured over the entire waveform Peak To Peak: Peak-to-peak = Max –...
  • Page 38 Hantek 6022BE Period: Time to take for the first signal cycle to complete in the waveform Frequency: Reciprocal of the period of the first cycle in the waveform Rise Time: Time taken from lower threshold to upper threshold Fall Time: Time taken from upper threshold to lower threshold...
  • Page 39: The Utility Function

    Hantek 6022BE Note: The results of the automatic measurements will be displayed on the bottom of the screen. Maximum 8 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left, out of screen.
  • Page 40 Hantek 6022BE be recalled at anytime by user. The Factory Setup function does not reset the following settings:  Language option  Date and time 3.6.2 Language Click “Language” in “Utility” menu. There are four languages in “Language” menu. The default language is English.
  • Page 41 Hantek 6022BE The Step Interpolation: The Linear Interpolation: User’s Manual...
  • Page 42: The Display System

    Hantek 6022BE The Sin(x)/x Interpolation: Note: The default interpolation mode is Sin(x)/x. 3.7 The Display System Click “Display” in main menu. The following figure shows the display system menu. Displays parameters setting.  Type  Grid  Intensity 3.7.1 Type Click “Type”...
  • Page 43 Hantek 6022BE If the Vectors type mode is selected, the waveform will be displayed as following figure. If the Dots type mode is selected, the waveform will be displayed as following figure. User’s Manual...
  • Page 44 Hantek 6022BE 3.7.2 Grid Click “Display” in main menu. The grid shows: User’s Manual...
  • Page 45 Hantek 6022BE Without the grid show: 3.7.3 Intensity Click “Display->Intensity” in main menu The following figure shows the intensity dialog. It shows the display parameters setting. You can change the grid and waveform color intensity in this dialog. User’s Manual...
  • Page 46: Save/Load Menu

    Hantek 6022BE 3.8 Save/Load Menu Click “File” in main menu to save waveform, setups and screen Save Data Save waveform data Save Setups Save the current oscilloscope setups to file Load Menu Click “File” in main menu to recall saved waveform, setup...
  • Page 47: Print

    Hantek 6022BE 3.9 Print Click “Print” in File. 1. Click “Print” in “File” menu to set the printer to print the current waveform. 2. Click the “PrintPreview” in “File” menu to get into the Preview window. In “PrintPreview” window, use the button to change the size of the waveform graph.
  • Page 48: Chapter 4 Application Example

    Hantek 6022BE Chapter 4 Application Example  Simple Measurement  Capturing a Single-Shot Signal  The Application of the X-Y  Taking Cursor Measurement User’s Manual...
  • Page 49: Simple Measurement

    Hantek 6022BE 4.1 Simple Measurement To acquire and display a signal, please do the steps as follows: Connect signal to CH1 by using probe The oscilloscope set the vertical, horizontal, and triggers controls at the best status automatically. Also, you can adjust the controls to meet your measurement to optimize the waveform display.
  • Page 50: Capturing A Single-Shot Signal

    Hantek 6022BE The results of measure display in Information output. 4.2 Capturing a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowledge of the signal in order to set up the trigger level and slope correctly. For example, if the event is derived from 3.3V COMS logic, a trigger level of 1.2 or higher Volts should work on a rising edge.
  • Page 51: The Application Of The X-Y

    Hantek 6022BE Drag the trigger level sign on the waveform display screen to proper position. It’s usually higher a little above the normal level. When the trigger conditions are met, data appears on the display representing the data points that the oscilloscope obtained with one acquisition.
  • Page 52 Hantek 6022BE Instruction of the Ellipse Method Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals. From the formula above: θ = ±arcsine (A/B) or ±arcsine (C/D) θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I and III quadrant, .
  • Page 53: Taking Cursor Measurements

    Hantek 6022BE 4.4 Taking Cursor Measurements Use cursors to make time and amplitude measurements on a waveform quickly. Measure the Peak Frequency or Time of the First Sine Waveform Do these steps: Click “Cursor->Source”, select CH1 (select CH2 if you want measure CH2).
  • Page 54 Hantek 6022BE Note: Click “Cursor->Type”, select “Cross”, you can measure time and amplitude at one time. User’s Manual...
  • Page 55: Appendix

    Hantek 6022BE Appendix  Appendix A: Specifications  Appendix B: General Maintenance User’s Manual...
  • Page 56: Appendix A: Specification

    Hantek 6022BE Appendix A: Specification Specifications Table: Input Max. sample rate Real-time sampling:48MSa/s Channels 2 Channels 20MHz (-3dB) Bandwidth Vertical resolution 8 bits/channel 20mV~5V/div@x1 probe(20mV,50mV,100mV,200mV, 500mV,1V,2V,5V/div 1,2,5 sequence) Gain range 200mV~50V/div@x10 probe 2V~500V/div@x100 probe 20V~5KV/div@x1000 probe Range 8 divisions Offset level...
  • Page 57 Hantek 6022BE width, rise time (10%~90%), fall time (10%~90%), positive duty cycle, negative duty cycle Math Addition, Subtraction, Multiplication, Division Rectangular, Hanning, Hamming, Blackman Window Physical Interface Universal Serial Bus(USB 2.0) No external power source required. Power Bus-powered from USB...
  • Page 58: Appendix B: General Maintenance

    Hantek 6022BE Appendix B: General Maintenance General Care Do not store or leave the oscilloscope where the device will be exposed to direct sunlight for long periods of time. Caution To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents.

Table of Contents