5-4-3V OP-05 voltage (0 10V) Analog out
* The voltage output occurs proportionally the voltage according to the size of a weight
In 0V ~10V.
* The type of voltage output can be changed according to SET UP F60
▩ SPECIFICATION
output Voltage
Precision
Min Impedence
▩ 9P D-TYPE Female & Voltage out circuit
1
2
※NOTICE
As NO 5 Terminal is not GND, Please don`t connect the device such as GND line or BODY GND
▩ Adjustment
How to calibrate for output rate bewteen 0v and 10v.
* The voltage out is to 0V when the weight is displayed 0 kg in indicator.
* The voltage out is to 10V when the weight is displayed max.capacity in indicator.
* If analog output is not correct,
You can make a fine adjustment with VR1(Zero adjustment) and VR2(Span adjustment)
on analog pc board by multi meter.( Recommended accuracy : 1/3,000 )
---------------------------------------------------------------------------------
FINE MECHATRONICS
0 10V DC out
Max 1/3000
Over 1 kΩ
1
3
6
2
7
3
8
4
9
5
FS8000
(+) 1
(-)5
1 FG
2 RXD(+)
3 RXD(-)
6 TXD(+)
7 TXD(-)
47