FINE mechatronics FS-8000 Instruction Manual page 22

Digital weighing indicator
Table of Contents

Advertisement

☞ 4Step
Indicator will display the capacity at weight column which was set at 2 step
after being displayed " SPAn ".
Please input the value of standard weight for span adjustment by numeric key.
This value of span standard weight must be equal to full capacity,or over 10% of
full capacity.
( In case of less 1/5,000 resolution ,the value of standard weight must be
over 10% of full capacity at least.)
( In case of over 1/5,000 resolution ,the value of standard weight must be
over 20% of full capacity at least.)
(Notice) If span capacity is set less 10% ,indicator will display error message.
(E r r 02 or E r r 03)
☞ 5Step
Please put the span standard weight on the platform.(the weight is 1000kg at here)
Press SET key after stable of platform.
CAL
(Notice) If indicator is unmatched with load cell capacity or span standard weight,
indicator will display error message (E r r 04)
☞ 6Step
Indicator will display any constant value of span adjustment.
If the range of this constant value is between 0.5000 -- 3.50000,
All procedure of span adjustment is normal.
And then,press SET key for next procedure.
If you remember this constant value , you can adjust the span without standard
weight by F99 (Function number 99 ) at set-up mode.
(Please remember this constant value,full capacity and one digit for
your further calibration & reference )
☞ 7Step
The "END" message is displayed in 6 step,
all span adjustment is end.
Press SET key after put down of span standard weight on the platform.
CAL
The indicator will enter into user's weighing mode.
---------------------------------------------------------------------------------
FINE MECHATRONICS
CAL
FS8000
22