3-3. ERROR MESSAGES & ADJUST
tESt or FS-XXXX
ERR--01
①Cause : In case resolution (A Interval/Max.display weight) was set over 1/20,000 resolution.
②Adjust : Set under 1/20,000 resolution(A Interval/Max.display weight)
ERR--02
①Cause : In case Standard Balance weight was more than Max CAPACITY
②Adjust : Make Set Standard Balance weight equal or less than Max CAPACITY
ERR--03
①Cause : In case Standard Balance weight for span adjust was set less than 5% of Max CAPACITY
②Adjust : Set Standard Balance weight for span adjust into less than 5% of Max CAPACITY
ERR--04
①Cause : In case the weight was not safty when it account the value of a span constant
②Adjust : Adjust a span after removing a cause to be unsafty
ERR--05
①Cause : In case the acual weight was more than Standard Balance weight
Or the amplification quanity of Analog circuit inside was more than.
②Adjust :
- Please check it if the actual weight was more than Standard Balance weight or not
If it did so,please adjust the standard weight into the value set up.
- If it continue to display ERR--05,please adjust NO 7,8 of Dip-Switch on the rear panel.
For a reference,
NO 1 ~6 of Dip-Switch is to adjust ZERO.
NO 7 ~8 of Dip-Switch is to adjust SPAN
Also because ZERO was changed according to NO 7 ~8 of Dip-Switch,
Please adjust ZERO again as
---------------------------------------------------------------------------------
FINE MECHATRONICS
: If indicator display only " tESt " or FS-XXXX (Model number)
without any operation ,first of all you must adjust "dip switch"
of back side panel for span and zero value.
3-1 ZERO Adjustment
FS8000
24