- The way to use the Dip-Switch.
NO 7
ON
OFF
ON
OFF
- Please adjust SPAN again after adjusting less than the present adjusted value.
※ If it continue to display ERR--05 in spite of adjusting the Dip-Switch as the above,
Please check it if the cable wire of a Load cell was normal or nor.
ERR--55
※
①Cause : In case a cable wire of a Load cell was connected on reverse.
②Adjust : Please check the connection of a Load cell as a reference of 2.3 CHAPTER
ERR--06
※
①Cause : In case the actual weight was loaded less than standard balance weight
Or was less than Analog Circuit Amplification.
②Adjust :
- Please adjust a standard balance weight into the weight set up.
- If continue to display ERR--06,Adjust NO 7,8 of the Dip-Switch on the rear panel.
For a reference,
NO 1 ~6 of Dip-Switch is to adjust ZERO.
NO 7 ~8 of Dip-Switch is to adjust SPAN
Also because ZERO was changed according to NO 7 ~8 of Dip-Switch,
Please adjust ZERO again as
- The way to use the Dip-Switch.
- Please adjust SPAN again after adjusting less than the present adjusted value.
NO 7
ON
OFF
ON
OFF
※ If it continue to display ERR--06 in spite of adjusting the Dip-Switch as the above,
Please check it if the cable wire of a Load cell was normal or nor.
---------------------------------------------------------------------------------
FINE MECHATRONICS
NO 8
ON
ON
OFF
OFF
3-1 ZERO Adjustment
NO 8
ON
ON
OFF
OFF
FS8000
Amplification Size
Small
1time
Normal
2times
Big
3times
Very Big
4times
Amplification Size
Small
1time
Normal
2times
Big
3times
Very Big
4times
OFF
ON
OFF
ON
25