Output Waveform Rise And Fall Time; Frequency Verification - Wavetek 75 Instruction Manual

Arbitrary waveform
Table of Contents

Advertisement

(REAR PANEL)
MODEL 75
FREQUENCY COUNTER
Figure 5-4.
Frequency Verification Equipment Connection
Table 5-5.
Frequency Verification
Table 5-6.
Frequency Verification Limits
PresslAdjust
1 . Setup, RSET (F2), EXEC (F3)
2. Parameter, SAMPL (F2), Control knob
Remarks
Sets default values.
Set the sample frequency to each of the settings
given in table
5-6.
Verify on the frequency counter
that each frequency setting falls within the upper
and lower limits given in the table.
NOTE: Set the frequency counter to measure period.
SETTING
1.0000
MHz
500.0
kHz
300.0
kHz
100.0
kHz
10.000
kHz
1.000
kHz
100.0
Hz
5.8
OUTPUT WAVEFORM RISE AND FALL TIME
This paragraph verifies that the rise and fall time of the
Model
75
output waveform falls within the specified
limits. If the readings are outside the specified limits,
perform the steps given in paragraph
5.4
(Calibration)
and then recheck using this paragraph. If the instrument
still fails to fall within specifications, refer to Section
6
(Troub1eshooting)and Section 4(Circuit Description)for
help in solving the out of tolerance problem.
UPPER LIMIT
1.0001 50
MHz
500.050
kHz
300.045
kHz
100.01 50
kHz
10.001 50
kHz
1.0001 50
kHz
100.01 50
Hz
1 . Interconnect the Model 75 and test equipment as
shown in figure
5-5.
Use the
50R
termination with
the carbon element in this step.
2. Set the osc~lloscope controls as follows:
Vertical Input
CHAN 1
Vertical Gain
2VIDIV
TIMEBASE
0.2
pslDlV
Trig Source
External
3.
Perform the steps in the order shown in table
5-7.
LOWER LIMIT
0.999850
MHz
499.925
kHz
299.955
kHz
99.9850
kHz
9.99850
kHz
0.999850
kHz
99.9850
Hz