Wavetek 75 Instruction Manual page 52

Arbitrary waveform
Table of Contents

Advertisement

Table 3-2.
Programming Reference Guide (Continued)
I
Scans by ArtekMedia O 2006
Function
Terminator
Select
Function
Mode
Trigger
Slope
ASCll
Character
XV
C
B
Q*
Description
When the power is turned on, the terminator or End Of String (EOS) character
(see EOI description, paragraph 3.28.1) is the ASCII line feed 'LF'. If the GPlB
controller does not look for the end message (EOI line low), and it does not
recognize the line feed (LF) as a string terminator, a new EOS character will be
needed. The EOS terminator may be changed to any ASCII character (except IVUL)
by sending 'XV' followed by the decimal value (of the ASCII code) corresponding
to the new terminator.
The character 'C' followed by a single digit code between 0 and 8selects standard
waveform which is inserted between start and stop addresses.
FUNCTION
CODE
Sine wave
co
Triangle wave
C1
Square wave
C2
Ramp Up
C3
DC
C4
Ramp Down
C5
Cosine
C6
Inv Sine
C7
Haver Sine
C8
The character 'B' followed by a single digit code between 0 and 4 selects the
operating mode of the instrument.This is the same as the MODE soft key.
MODE
Continuous
,ODT
BO
Triggered
B1
Gated
B2
Toggled
B3
Burst
B4
The character 'Q' selects generator triggering or gating on the rising or falling
edge of the Trig In ( T L ) signal or GPIB 'J' (positive slope) or 'H' (negative slope)
command. See TRlGGERlNGlGATlNG CHART below.
TRlGGERlNGlGATlNG CHART
Positive Slope
(Rising Edge) QO
'J'
Trig: Single cycle
J
Gate: Free runs
Toggle: Toggles
Burst: Start burst
'H'
Trig: Enable 'J'
Gate: Off
Toggle: Enable 'J'
Burst: Enable 'J'
Negative Slope
(Falling Edge) Q1
Trig: Enable 'H'
Gate: Off
Toggle: Enable 'H'
Burst: Enable 'H'
Trig: Single cycle
Gate: Free runs
Toggle: Toggles
Burst: Starts Burst