Cleaning and maintenance
52/68
–
Only touch new lamps in the handling zone.
–
Test the function of the safety switch.
The safety switch ensures that the device switches off in the event of a
faulty filter panel.
–
Press the switch.
If the switch springs back automatically then it is functional.
–
Replace the filter panel.
4.9
Cleaning or replacing UV lamps in the
base
–
Lift up the acrylic glass panel and then remove the lamp covers and the
partition: see page 43.
Replacing lamps: see page 49.
The assembly is the same process in reverse order.
When assembling the device please note:
–
Clean the partition.
–
Hold the partition with a cloth in order to prevent finger prints.