Cleaning and maintenance
58/68
–
Screw the lid with the hose onto the new canister.
–
Push the hose well down so that it touches the bottom of the canister.
–
Put the canister back in the device.
–
Close the bed base.
Venting
NOTE:
Ventilate after changing the canister by switching over to
preset mode and then select Function 231 (see 'Presettings',
order no. 1011777-..).
–
Press the AROMA button.
–
Close the canopy.
The system will now be ventilated for 1 minute. AQUA SYSTEM liquid will be
released from the nozzles during this period.
–
Test the AQUA SYSTEM function afterwards. Repeat the ventilation
process if it does not work.