07/2015
1015722-00B
–
Only touch new lamps in the handling zone.
–
Test the function of the safety switch.
The safety switch ensures that the device switches off in the event of a
faulty filter panel.
–
Press the switch.
If the switch springs back automatically then it is functional.
–
Replace the filter panel.
4.9
Cleaning or replacing
UV lamps in the base
–
Lift up the acrylic glass panel and the partition: see page 48.
–
Remove the lamp covers at the head and foot ends.
Replacing lamps: see page 54.
The assembly is the same process in reverse order.
Cleaning and maintenance
Operating instructions – 57/84