6 Electromagnetic compatibility
6.3 Working clearances
Rated maximum output power of
transmitter
[W]
0.01
0.1
1
10
100
20
Working clearance according to transmission frequency [m]
150 kHz to 80 MHz
d= [1.2] √P
0.12
0.38
1.2
3.8
12
For transmitters whose maximum nominal output is not specified in the
above table, the recommended working clearance d in meters (m) can be
determined using the equation in the corresponding column, where P is
the maximum nominal output of the transmitter in watts (W) specified by
the manufacturer of the.
Remark 1
The higher frequency range applies at 80 MHz and 800 MHz.
Remark 2
These guidelines may not be applicable in all cases. The propagation of
electromagnetic waves is influenced by their absorption and reflection by
buildings, objects and persons.
Operating Instructions SIROInspect
80 MHz to 800 MHz
d= [1.2] √P
0.12
0.38
1.2
3.8
12
Sirona Dental Systems GmbH
800 MHz to 2.5 GHz
d= [2.3] √P
0.23
0.73
2.3
7.3
23
63 82 936 D3584
D3584.201.01.03.02 02.2013