Common Specifications - Fuji Electric FD5000 series Instruction Manual

Table of Contents

Advertisement

Instruction Manual for FD5000 Series
8.1.14 Temperature Measuring Unit (RTD)
Input
Range
Indication
sensor
PA
PT1001
-100.0 to 199.9°C
(-148.0 to 391.8°F)
JPA
JPt1001
PB
PT1001
-100 to 600°C
(-148 to 1112°F)
JPB
JPt1001
Input circuit : Single ended type
Operating system : 6 Y conversion
Maximum sampling rate : 12.5 times per second
Current through RTD : About 1 mA
External resistance : 101 or less per wire
Linearizer : Digital linearizer
Burnout alarm : It blinks by ---- display.
8.1.15 Frequency Measuring Unit (open collector, logic, and magnet)
Measurement
Range
Indication
range
11
0.1 to 200Hz
12
1 to 2000Hz
Pre-scale : 0.001 to 5
Frequency division : 1 to 100
13
0.01 to 20kHz
14
0.1 to 200kHz
Input type
Input voltage lebel
Open collector
LO : 1V or less (5V : 2.2 k 1 pull up)
Logic
LO : 1V or less,HI : 2.5 to 15V
Magnet
0.3 to 30Vp-p
Duty ratio : 50%
8.1.16 Frequency Measuring Unit (500 Vrms)
Measurement
Range
Indication
range
11
0.1 to 200Hz
12
1 to 2000Hz
Pre-scale : 0.001 to 5
Frequency division : 1 to 100
13
0.01 to 20kHz
14
0.1 to 200kHz
Input type
Input voltage lebel
Voltage
50 to 500Vrms
Duty ratio : 50%
8.1.17 Strain Gage Unit
Sensor
Zero adjusting range Span adjusting range
power
5V
-0.3 to +2mV/V
1 to 3mV/V
10V
Input circuit : Single ended type
Operating system : 6 Y conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
8.1.18 Process Signal Measuring Unit
Measurement
Range
Indication
range
1V
1 to 5V
Offset : ± 9999
Full scale : 0 to ± 9999
2A
4 to 20mA
Input circuit : Single ended type
Operating system : 6 Y conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Highest
Accuracy
resolution
0.1°C
± (0.15% of FS)
(0.1°F)
1°C
± (0.3% of FS)
(1°F)
Renewal
Highest
time of a
Accuracy
resolution
display
0.1Hz
0 to 10s
1Hz
1s
±(0.2% of FS)
10Hz
100ms
100Hz
Maximum
permissible input
30V
15V
Renewal
Highest
time of a
Accuracy
resolution
display
0.1Hz
0 to 10s
1Hz
1s
± (0.2% of FS)
10Hz
100ms
100Hz
Maximum
permissible input
500Vrms
Highest
Accuracy
resolution
0.5 V/digit
± (0.1% of FS +2digit)
1 V/digit
Maximum
Input
permissible
Accuracy
impedance
input
About
± 100V
100M1
± (0.2% of FS)
About
± 100mA
101

8.2 Common Specifications

Display :
7-segment LED display (character height : 14.2 mm on main
display and 8 mm on sub-display)
Automatically indicated when the calculated result is
Polarity indication :
negative.
-9999 to 9999
Indication range :
OL or -OL for input signals outside the indication range
Over-range alarm :
Decimal point : Can be set at an arbitrary digit.
Zero indication :
Leading zero suppression
External control :
HOLD, PH, DZ (reset for frequency measuring unit)
Operating temperature and humidity range : 0 to 50 °C, 35 to 83% RH
Storage temperature and humidity range :
100 to 240 V AC ± 10% for AC power supply unit
Power supply :
9 to 50 V DC for DC power supply unit
Power consumption :
Approx. 5 W
96 mm (W) x 48 mm (H) x 146.5 mm (D)
External dimensions :
Note : Depth (D) denotes the maximum value.
Weight :
450g
Withstand voltage :
2000 V AC for 1 min. between power terminals and
input terminal, and between power terminals and each
output terminal (AC power supply)
Withstand voltage
500 V DC for 1 min. between power terminals and input
:
terminal, and between power terminals and each output
terminal (DC power supply)
Withstand voltage
500 V DC for 1 min. between input terminal and each
:
output terminal, and between analog output terminal and
communication terminals
2000 VAC for 1 min between case and each output
terminal (common to both AC and DC supply)
Insulation
100 M1 between the above terminals when 500 V DC
:
resistance
is applied
8.3 Output Specifications
8.3.1 Output for Comparison
Conditions for comparison
Indicated value > Upper limit judgment value
Lower limit judgment value ) Indicated value ) Upper limit judgment value
Lower limit judgment value > Indicated value
Control system : Micro computer operating system
Judgment value setup range : -9999 to +9999
Hysteresis : Can be set in the range of 1 to 999 digits for each judgment
value
Operating speed : Depends on the sampling rate.
Output method : Relay contact output (Make and break contacts for HI and
LO and make contacts for GO)
Output rating : 240 V AC, 8 A (resistive load) and 30 V DC, 8 A (resistive
load)
10/12
(non-condensing)
-10 to 70 °C, 60% RH or less
Judgment
result
HI
GO
LO

Hide quick links:

Advertisement

Table of Contents