Terminal Arrangement; External Controls - Fuji Electric FD5000 series Instruction Manual

Table of Contents

Advertisement

Instruction Manual for FD5000 Series
Caution
!
(1) Mount the product to a panel that is strong enough to hold the
product. If the panel is not strong enough or the product is not fixed
tightly, it may fall down and cause injury.
The A5000 does not have a power switch, and will thus be
(2)
immediately ready for operation upon connecting it to a power
supply.
(3) If the product is installed inside other equipment, provide sufficient
heat dissipation to ensure that the temperature inside the
equipment does not exceed 50°C .
3.

Terminal Arrangement

Comparison
Serial
output
communication
12 13 14 15 16 17 18 19
Input (the shape depends on the unit.)
3.1
Power
No.
10 POWER
10 11
11 POWER
3.2

External Controls

No.
6 7
8 9
6
7
8
9
3.3
Input Signals
3.3.1 DC Voltage Measuring Unit (Range 11)
No.
1
2
1 2
3
3
3.3.2 DC Voltage Measuring Unit (Range 12)
No.
1
2
1 2
3 4
5
3
4
5
Analog output
24 25 26
20
22
21
23
6
7
8
9
10 11
External
Power
control
connector
Name
Description
Power terminal without polarity for both DC
and AC
Power terminal without polarity for both DC
and AC
Name
Description
Control for hold function. Enabled when
HOLD
short-circuited or at the same potential as
COM.
Control for digital zero function. Enabled
DZ
when short - circuited or at the same
potential as COM.
Control for peak hold function. Enabled
PH
when short - circuited or at the same
potential as COM.
COM
Common for all external control terminals.
Name
Description
HI
Positive input terminal
NC
Do not connect this terminal.
LO
Negative input terminal
Name
Description
Positive input terminal for range 12
12
(± 999.9 mV)
Positive input terminal for range 13
13
(± 9.999 V)
Positive input terminal for range 14
14
(± 99.99 V)
Positive input terminal for range 15
15
(± 600 V)
LO
Negative input terminal
3.3.3 DC Current Measuring Unit
No.
1
2
1 2
3 4
5
3
4
5
3.3.4 AC Voltage Measuring Unit (Ranges 11 to 13)
No.
1
1 2
3
2
3
3.3.5 AC Voltage Measuring Unit (Ranges 14 and 15)
No.
1
2
1 2
3
3
3.3.6 AC Current Measuring Unit (Ranges 23 to 25)
No.
1
2
1 2
3 4
5
3
4
5
3.3.7 AC Current Measuring Unit (Range 26)
No.
1
2
1
2
Applicable solderless terminals
3.3.8 Resistance Measuring Unit
No.
1
2
1 2
3 4
5
3
4
5
3.3.9 Temperature Measuring Unit (TC)
No.
1
1 2
3
2
3
Name
Description
Positive input terminal for range 23
23
(± 9.999 mA)
Positive input terminal for range 24
24
(± 99.99 mA)
Positive input terminal for range 25
25
(± 999.9 mA)
LO
Negative input terminal
Name
Description
Positive input terminal for ranges
11-12
11 (99.99 mV) and 12 (999.9 mV)
Positive input terminal for range
13
13 (9.999 V)
LO
Common input terminal
Name
Description
Positive input terminal for range
14
14 (99.99 V)
Positive input terminal for range
15
15 (600 V)
LO
Common input terminal
Name
Description
Positive input terminal for range
23
23 (9.999 mA)
Positive input terminal for range
24
24 (99.99 mA)
Positive input terminal for range
25
25 (999.9 mA)
LO
Negative input terminal
Name
Description
HI
Input terminal
LO
Input terminal
.
5.8 mm
min
Name
Description
HI
Input terminal for all ranges
LO
Input terminal for all ranges
Constant current for four-wire resistance
+S
measurement (positive)
Constant current for four-wire resistance
-S
measurement (negative)
Common terminal (grounding terminal for
LO
input circuit)
Name
Description
+
Positive terminal for thermocouple
NC
Do not connect this terminal.
-
Negative terminal for thermocouple
2/12
.
5.8 mm
min

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents