7.5 Running the Test

There are five command buttons in the Control Panel:
• Start: This button begins a timed test.
• Clear Display: Clears the statistical data, on the lower left side of the
screen, and the graph, on the right side of the screen.
• Save to Disk: Saves the current data to a file in the computer.
• Configure: Returns to the megohmmeter Setup dialog box for configur-
ing the instrument.
• Download: Retrieves data from the memory of the megohmmeter for
storage in the computer.
To run a timed test, perform the following steps:
1.
Select the appropriate voltage for the test by clicking on one of the six radio
buttons next to the desired selection. If Var V or Var Ramp is selected, you
will also need to select profile 1, 2 or 3 using the drop-down arrow next to that
selection.
2. Press the Start button. This will run the test at the selected voltage. As the
test is running, all other buttons are grayed out and the Start button becomes
a Stop button.
Once the test has started, a graph will begin to appear and build on the right side
of the Control Panel representing a standard test and a step voltage test respec-
tively.
Megohmmeter Model 5070
Figure 7-9
79