4.2 MODE / PRINT Button

4.2.1

Primary Function - MODE

The primary function of this button is used before the measurement takes place to
define the measurement conditions.
NOTE: This button is inactive in the Step Voltage and SET-UP positions.
• Press the MODE button once to access the list of possible measurement modes.
Select the mode using the ▲ ▼ and ◄ ► buttons.
• To validate the mode selected, press the MODE button again.
The measurement mode choices are as follows:
MANUAL STOP
This is the conventional quantitative
insulation measurement mode.
The measurement is started by press-
ing START/STOP button and stopped by
pressing START/STOP button again.
In this mode, the user determines the
duration of the test, which will be indi-
cated by the elapsed time indicator.
MANUAL STOP + DD
The measurement is started by press-
ing START/STOP button and stopped by
pressing START/STOP again.
1 minute after the end of the measure-
ment, the instrument calculates and dis-
plays the Dielectric Discharge (DD) ratio.
The time remaining during this minute is
displayed.
TIMED RUN
This mode is used to perform a measure-
ment for a duration defined in advance,
with a predetermined number of mea-
surement samples. The measurement
is started by pressing the START/STOP
button and stops automatically after
the time programmed by the user has
occurred.
28
MODE
Total Run Time
Manual Stop
Manual Stop + DD
Timed Run
Timed Run + DD
DAR (s/s)
PI (m/m)
MODE
Total Run Time
Manual Stop
Manual Stop + DD
Timed Run
Timed Run + DD
DAR (s/s)
PI (m/m)
MODE
Total Run Time
Manual Stop
Manual Stop + DD
Timed Run
Timed Run + DD
DAR (s/s)
PI (m/m)
Megohmmeter Model 5070
Duration Sample
(h:m)
(m:s)
02:30
01:40
30/60
1/10
Duration Sample
(h:m)
(m:s)
02:30
01:40
30/60
1/10
02:30:00
Duration Sample
(h:m)
(m:s)
02:30
01:40
30/60
1/10