rent, and, possibly, DAR, PI, DD (if available), temperature corrected resistance
and the R(t) graph.
WARNING: If a button other than MEM is pressed, or the selector switch
is moved before pressing MEM a second time, the instrument exits from
MEM mode without saving the results.
Memory Capacity
• Total memory space: 128kb
• Data management: 8kb
• Free memory space: 120kb
An insulation measurement requires 80 bytes. Therefore, approximately 1500
insulation measurements can be stored.
Free Memory Space
This function is automatically activated when a result is saved.
Press the MEM button once to bring up the first available Object and Test storage
location. A bargraph is displayed corresponding to the available memory location.
• If the entire memory is free, all of the bargraph segments appear.
• If the entire memory is full, the arrow to the left of the bargraph flashes.
• As soon as the storage is complete, the bargraph disappears.
Each segment of the bargraph equals approximately 50 recordings.
4.8.2

Second Function - MR (Recall)

The MR function allows you to recall any of the stored data from memory, regard-
less of the position of the rotary selector switch, except OFF or SETUP.
When the MR button is activated, the following screen is displayed:
• The flashing cursor identifies the last occupied Obj. and Test number.
Here - 47 : 99
• Use the ▲ , ▼ , ◄ , ► buttons to select
the desired Obj. and Test number.
• After selecting the Obj. and Test, press
the ► button to access the first item of
information relative to this measure-
ment.
• To access more data, press DISPLAY
repeatedly, or GRAPH, if the mode
selected before the start of the
measurement allows.
To exit the MR function, press the MR button once again or turn the selector
switch to another position.
Megohmmeter Model 5070
Recall
MEMORY
Obj. Test
Date
47 99 12.15.2003 07:04 625V
13 59 12.07.2003 22:39 3800V
13 58 11.29.2003 15:47
02 03 11.24.2003 15:04 2150V
02 02 10.29.2003 21:45 975V
02 01 09.30.2003 02:43 5000V
01 02 09.02.2003 15:07
Time
Fct.