8.2 Fuse Replacement

If GUARD FUSE appears on the display, the fuse must be replaced, taking all the
necessary precautions when opening up the instrument.
WARNING: Make sure that none of the terminals are connected and that
the selector switch is set to the OFF position before opening up the case.
Only use the type of fuse shown on label inside the unit's cover:
0.1A - Fast Acting - 380V, 5x20mm, 10kA (Cat. #2119.84)

8.3 Cleaning

WARNING: Disconnect the instrument from any source of electricity.
• Use a soft cloth lightly dampened with soapy water. Rinse with a damp
cloth and then dry with a dry cloth.
• Do not get water inside the case. This may lead to electrical shock or
damage to the instrument.
• Do not use alcohol, solvents or hydrocarbons.

8.4 Storage

If the instrument is not used for an extended time period (longer than two months),
it is recommended that a complete charge and discharge of the battery is per-
formed three times before re-using the instrument.
Complete battery discharge can be performed:
• By removing the battery from the instrument and applying a 3 Amp load
to it.
• On the 5000V position with the backlight on (consumes the most power)
86
Megohmmeter Model 5070