Connecting the Balance to AC Power
When your balance has reached room temperature, simply plug
the AC adapter into the rear of the balance and plug into an appro-
priate AC outlet. The balance will turn on automatically. For best
results, the balance should be left plugged in at all times.
USB
Connection
Interface

Leveling your Balance

It is necessary to level the balance whenever the balance is moved.
Do not turn the balance over.
The number of feet varies on each model:
Analytical models: 2 adjustable front feet, one fixed rear foot.
Toploading models with round pan: 2 front feet, one rear fixed foot.
Toploading models with square pan: 4 adjustable feet, one in each
corner.
To level your balance:
1. Turn all leveling feet counterclockwise until the feet are fully
retracted into the base of the balance.
2. Note the location of the bubble in relation to the circle. For maxi-
mum weighing accuracy, the bubble should be located inside the
black ring.
3. Begin with the foot that is opposite of the location of the bubble
and turn clockwise until the bubble is moved into the black ring.
If necessary, repeat this step with the other leveling feet until the
bubble is positioned in the center of the black ring.
4. Make sure that all feet are touching the countertop and no rock-
ing occurs when the corners of the balance are pressed.
The bubble moves TOWARDS a foot when that foot is
turned CLOCKWISE. The bubble moves AWAY from a foot
Note:
Memo:
when that foot is turned COUNTERCLOCKWISE.
RS232
Ethernet
Connection
5
Lock-Down
Power
for cable
and lock