Environmental Settings - Denver Instrument P-114 Operation Manual

Pinnacle series
Hide thumbs Also See for P-114:
Table of Contents

Advertisement

Environmental Settings

Your balance can be set up for optimized weighing to compensate
for varying conditions including vibration and drafts.
Adjusting Filter Settings
The filter setting is set to take a number of readings and display the
average on the screen. If you are in a very stable environment (i.e.
on a marble slab and away from the heating/cooling system) there
is no need to perform a high amount of filtering this also allows for
a quicker response time. On the contrary, if the balance is in an
unstable environment (i.e. on a production floor) quite a bit of fil-
tering may be needed to give good results.
1. Press MENU.
2. Press the
until "Environment" is highlighted, press the
3. Press the
until "Filter" is highlighted, press the
4. Press
or
arrow until the desired filter setting is highlighted
(low, normal or high) and press ENTER.
5. Press the TARE key to return to the measurement screen, or
return to the main menu..
Applications
Units
Environment
GLP/ISO
Print
Language
System
Standby
Filter
Stability
Auto Zero
System Screen
28
.
.
to

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents