Calibration - Denver Instrument P-114 Operation Manual

Pinnacle series
Hide thumbs Also See for P-114:
Table of Contents

Advertisement

Primary
Secondary
Toggling between Primary and Secondary Unit
Press ENTER to exchange the primary and secondary units while on
the measurement screen.

Calibration

Your balance was calibrated at the factory; however, it is necessary to
re-calibrate upon setup and on a regular basis thereafter. The factory
recommendation for calibration is once per week using a permissible
weight standard. Reasons for more frequent calibration include:
• Moving the balance
• Significant temperature changes
• Removing the balance from AC power
• Procedures to meet ISO/GLP procedures
Grams
Kilograms
Milligrams
Ounces
Troy ounces
Pounds
Grains
more