To quit Formulation mode, select "Normal Mode" in applica-
tions menu.
Note:
Memo:

Animal Weighing

Animal weigh mode makes it easy to weigh animals that are contin-
uously moving as the weight is taken. This feature can also be used
for measurements taken in environments with extreme vibrations
and/or drafts.
There are two start settings.
Automatic - the count begins automatically when a weight is placed
on the pan.
Manual - press the enter key to begin the countdown.
You can also define the number of points to average (5, 10, 15 or
20). The higher the number of points selected, the more accurate the
results but the test time increases.
Entering animal weighing mode:
1. Press MENU.
2. When "Applications" is highlighted, press the
3. Press
until "Animal Mode" is highlighted and then press
4. Press the
or
5. Press
or
to select number of point to average.
5. Press the ENTER key to return to the measurement screen or
return to the main menu.
Beginning animal weighing:
1. "Animal" should be displayed on screen. If not follow steps to
"Entering animal weighing mode".
2. Remove all items from the pan, except a container if necessary.
3. Press the TARE key.
4. Place animal or sample on the balance.
5. Press ENTER (for manual start) and the balance will automatically
begin the countdown.
6. When the number of values to average is met, "LOCKED" will be
displayed along with the average value. A "L" will also appear as
the stability icon.
7. To Unlock the reading, remove the animal/sample from the pan or
press ENTER.
8. Repeat steps 2-7 for all samples.