Disclaimer
• Calibrate your balance using reference weights of the appropriate
tolerance (class). An instrument can be no more accurate than the
standard to which it has been compared. For assistance in the
selection of reference weights, please contact Denver Instrument
Company.
• Changes or modifications not expressly approved by the manu-
facturer could void the user's authority to operate this equipment.
• Use of this product in a manner not specified by the manufacturer
may impair any safety protection provided by the equipment.
• Do not drop objects on the pan of the balance.
• Never lift balance by the weighing pan as this may cause damage
to internal mechanism. Always lift and transport the balance by
its base, including removal from packing materials.
• If load exceeds 115% of maximum capacity, damage to the bal-
ance may occur.
Warnings and Safety
Information
Read the entire Operation Manual prior to attempting to operate
your precision balance. Connect only Denver Instrument acces-
sories as these are designed for optimal performance.
• Make sure the voltage rating printed on the AC adapter is identi-
cal to your local line voltage.
• Do not use this balance in a hazardous location.
• The only way to turn the power completely off is by disconnect-
ing the AC adapter from the balance.
• Protect the AC adapter from contacts with liquids.
• This product intended for indoor use.
This unit has no user serviceable parts! Do not open the balance
housing, as this will void the manufacturer's warranty.
!
WARNING!