parameter up.
To activate the ALARM function before or during a test, press the 2nd + ALARM
buttons. The word ALARM will appear at the top right side of the display.
If the insulation resistance drops below the set value at any time during the test,
the ALARM
as long as the alarm condition exists.
The alarm function can be deactivated at any time by pressing the 2nd + ALARM
buttons again.

BUTTON

screen to the right.
To activate the display backlight, press the 2nd +
To deactivate, press the 2nd +

3.8 MEM / MR BUTTON

The instrument can record the results and store them along with addresses to
identify the corresponding object and test.
The object, indicated as Obj on the display, can represent a single piece of
equipment, a department, or an entire facility. The Obj can hold up to 99 tests that
If the Obj is a 3-phase motor, then each test number will represent a
e.g. -
Test 1 = Phase A to housing
Test 2 = Phase B to housing
Test 3 = Phase C to housing
and so on.