Gummel Plot
Range and Integration Time
Medium integration smoothes the plot at the ultra low levels. Make sure the range is set to AUTO or
LIMITED to lowest current range:
HRSMU: 10 pA
MPSMU: 1 nA
HPSMU: 1 nA
Maximum resolution
16 PLC