ZERO CHECK

Channel Setup
Delete SMU1 and SMU2
Click the Classic Test tab, I/V Sweep, and Select to display the Channel Setup screen.
On this display, delete the rows of SMU1 and SMU2.
Change the Mode of SMU4 to V.
Change the Function of SMU4 to VAR1.
This setup will be used for checking the zero measurement on SMU4.
Change Mode to V and
Function to VAR1