5. Insert the flue gas probe through the probe mount into the flue
gas duct.
6.2.1.6.

Using the modular flue gas probe

Checking the thermocouple
The thermocouple of the flue gas probe must not lie against the
probe cage.
> Check before use. Bend the thermocouple back if necessary.
Aligning the flue gas probe
The flue gas must be able to flow freely past the thermocouple.
> Align the probe by turning it as required.
When using the testofix® probe mount, the measuring
instrument is exposed to heat radiating from the flue gas duct
during the measurement. In order to prevent consequently
affecting the combustion air temperature a AT probe with
cable must be used for the measurement!