3 Product description
3.2 Controls and functional elements of the DAC UNIVERSAL
32
3.2.5

Safety ring

The DAC UNIVERSAL is equipped with a safety ring that serves as a
cycle confirmation tool and prevents users from pinching their fingers
when closing the lid.
If there is an object between the chamber and the lid when the lid closes,
the cycle is stopped immediately. The lid is then lifted for 5 seconds and
a message is displayed.
To remedy message 86, see "Display messages DAC UNIVERSAL
Remedy display message_Sirona_Nitram
ADVANCED [ → 140]" or "Display messages DAC UNIVERSAL
STANDARD [ → 211]".
3.2.6

Water tank with sieve

The water tank is used for manually filling the DAC UNIVERSAL with
demineralized or distilled water (water conductivity < 3.0 µS/cm). The
sieve located on the water tank filters out any particles that may be
present in the water during filling.
NOTICE
The water should be poured carefully through the sieve into the tank.
Otherwise, the water may run in between the tank and the DAC
UNIVERSAL itself, resulting in water on the table or in worst-case, water
on the electrical parts of the DAC UNIVERSAL, which can cause a short
circuit.
IMPORTANT
The water must be free of foreign objects.