Code
CST
SCD
SAV <=(0 to 10)>
RCV
KLO
TCM<=(AUTO, MAN)>
TCS<=(-5 TO 105)>
RSR
RMD <=(PH, MV, TEMP)>
RMX <= -2 to 16 for pH,
±2000 for mV,
-5 to 105 for temperature>
RMN <= -2 to16 for pH,
±2000 for mV,
-5 to 105 for temperature>
RTX <=(-2132 TO 2132)>
RTN <=(-2132 to 2132)>
CMD <=(pH, mV)>
CSP <=(-2 to 16 for pH,
±2000 for mV)>
CDB <=(0 to 2 for pH,
0 to 100 for mV)>
CPR <=(0 to 100)>
CGN <=(0 to 99.99)>
CRS <=(0.00 to 99.99)>
CRT <=(0.00 to 99.99)>
CTX <=(-2132 to 2132)>
CTN =(-2132 to 2132)
A1M <= (PH HIGH, PH LOW,
MV HIGH, MV LOW,
TEMP HIGH, TEMP LOW)
A1S <=(-2 to 16 for pH,
±2000 for mV,
-5.0 to 105.0 for temp)>
A1H <=(0.01 to 2 for pH,
1 to 100 for mV,
0.1 to 2 for temp)>
A1D <=(0 to 30)>
A2M <=(PH HIGH, PH LOW,
MV HIGH, MV LOW,
TEMP HIGH, TEMP/LOW)>
Table 6 RS-232 Remote Commands
Function
GENERAL SETUP
Do a cold start
Set calibration defaults
Recall or set signal average
Recall current measurement values
Toggles keyboard lockout status and responds with the
new status
Recall or set temperature compensation mode
Recall or set temperature compensation setpoint
Reference solution timer reset
RECORDER
Recall or set recorder mode
Recall or set maximum value setpoint for current recorder
type
Recall or set minimum value setpoint for current
recorder type
Move recorder trim maximum count by indicated amount
Move recorder trim minimum count by indicated amount
CONTROLLER
Recall or set controller mode
Recall or set controller setpoint
Recall or set controller dead band
Recall or set controller preset
Recall or set controller gain
Recall or set controller reset value
Recall or set controller rate value
Move controller trim maximum count by indicated amount
Move controller trim minimum count by indicated amount
ALARMS
Recall or set alarm 1 mode
Recall or set alarm 1 setpoint for current alarm type
Recall or set alarm hysteresis for current alarm type
Recall or set alarm 1 delay time
Recall or set alarm 2 mode
47
Response
Prints setup
OK
Value (on recall); or OK (on
set)
Prints current values
Locked or Unlocked
Value (on recall); OK (on set)
Value (on recall); OK (on set)
OK
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
OK
OK
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
OK
OK
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)
Value (on recall); OK (on set)