Industrial Process Ph Electrode; Sample Handling System; Submersible; Flow-Thru Cell - Hach EC1000 Instruction Manual

Process ph/orp system
Table of Contents

Advertisement

Figure 4
Electrodes

1.6 Industrial Process pH Electrode

1.7 Sample Handling System

1.7.1 Submersible

1.7.2 Flow-Thru Cell

1.7.3 In-Line

The Hach Industrial pH Electrode's unique design does not use an electrolyte pump and
does not have a replaceable sensor. The low-maintenance electrode has self-pressurized
electrolyte and a flat electrode. The electrode features high temperature and pressure
resistance and is designed for traditional applications such as water and wastewater
analysis. The electrode is housed in a CPVC body with a built-in preamplifier and has the
same size and shape to fit all of the mounting configurations described in SECTION 5,
INSTALLATION.
Three sampling techniques can be used with the EC1000 pH/ORP System: submersible,
flow-thru cell, and in-line installations. Mounting kits are available for the different
applications. Refer to SECTION 5, INSTALLATION for detailed installation information.
For the submersible application, mount the electrode to the end of a pipe using the
Submersible Mounting Kit (Cat. No. 46401-00), see Figure 15 on page 61, and place it in
a tank or non-flowing sample.
The Flow-Thru Cell provides for atmospheric sampling of the process water, and easy
installation and removal of the electrode. Mounting Kit (Cat. No. 46403-00) is required,
see Figure 16 on page 61.
For the in-line application, place the electrode directly in-line with the sample through a
"Tee" pipe fitting using the In-line Mounting Kit (Cat. No. 46402-00), see Figure 15 on
page 61. Pressure reducers may be required if the sample pressure is greater than 35 psig.