2.2.2 Recorder Output

Press
. The DeLaY TIME is locked in, and the display returns to:
ENTER
Press
. The display backs out to:
CLEAR
Press the
key.
DOWN ARROW
The ALARM2 submenu is identical to ALARM1 submenu: mode; set point; hysteresis;
and delay time are all selected identical to ALARM1.
Press
and then
CLEAR
CLEAR
to exit the menu.
pH/mV
An isolated 4-20 mA recorder output is provided. The output also can be adjusted to
0-20 mA, and with the addition of a resistor across the recorder output lines, can be setup
as a 0 to 100 mV, 0 to 1 V, or 0 to 10 V voltage output. See Section 5.6.4 Recorder and
Controller for resistor installation. The recorder can be set to cover any portion of the pH,
mV, or temperature measurement range. The zero and full scale mA output can be
trimmed from the keyboard so the output can be calibrated to exactly 4-20 mA, or adjusted
to match the zero and full scale on the recorder being used.
To check or program the recorder settings, press the
SETUP menu.
Press the
key to scroll to:
DOWN ARROW
This option controls the minimum and maximum outputs shown on the
RECORDER submenu:
Press
.
ENTER
PH is the default. The mode options are: PH, MV, or TEMPerature. Select the mode by
pressing the
DOWN ARROW
displayed, press
.
ENTER
This option controls minimum and maximum settings on the recorder. Press
Press
again to display the current MINimum value.
ENTER
Press
and change the value by pressing the
ENTER
key or the
ARROW
UP ARROW
Press
to advance to:
ENTER
Press
and change the current MAXimum value by pressing the
ENTER
and the
key or the
DOWN ARROW
again to back out to the top level of the menu, or press
SETUP
key or the
key. When the desired MODE is
UP ARROW
RIGHT ARROW
key.
key.
UP ARROW
29
key to call up the
.
ENTER
key and the
DOWN
RIGHT ARROW
key