Changing Pacing Modes

8-6
6. Increase the output until cardiac capture occurs.
Press up
on
7. Decrease the output to the lowest level that still maintains capture.
Press
on
Output
Start
Press
to stop pacing. Press
Stop
green LED next to the button goes out, indicating pacing is no longer active.
Changing Pacing Modes
If paced pulses are being delivered, you must stop pacing before changing the
pacing mode:
Start
1. Press
to stop pacing.
Stop
2. Press
to change the mode.
Mode
3. Adjust the rate, if needed.
Start
4. Press
to resume pacing.
Stop
5. Adjust the output, as needed to obtain capture.

Defibrillating During Pacing

If the patient must be defibrillated during pacing, follow the procedure for
defibrillating in Manual Mode on page 6-4.
Pacing is automatically turned off when you charge the defibrillator and the
pacing dialogue box is removed from the display. After a shock, pacing
remains off.
To resume pacing, refer to "Pacing" on page 8-4. When you resume, the rate,
mode, and output settings start at the settings selected prior to defibrillation.
Output
to increase the output in increments of 10 mA.
to decrease the output in increments of 5 mA.
Pacer
to exit the pacing function. The
Pacing