Automatic Re-Analysis On; Automatic Re-Analysis Off - Philips HEARTSTART XLT M3500B Instructions For Use Manual

Defibrillator/monitor
Table of Contents

Advertisement

Defibrillating
Delivery of the shock is confirmed by the message
and the
shock counter is updated.
Figure 3-9 Shock Delivered Display
Shock Delivered
Shocks: 1
00:00:40

Automatic Re-analysis On

3
If Automatic Re-analysis is on, the HeartStart XLT analyzes the ECG fol-
lowing delivery of the shock. You are prompted to press
, if an addi-
tional shock is advised. This cycle repeats until the rhythm converts or a
shock series is complete. (A shock series may be configured to 2, 3, or 4
shocks.)

Automatic Re-analysis Off

If Automatic Re-analysis is off, the HeartStart XLT monitors the ECG for
potentially shockable rhythms (provided Rhythm Monitoring is on) and
prompts you to press
if one is detected. You can initiate analysis,
without being prompted, by pressing
.
M3500B HeartStart XLT Defibrillator/Monitor
3-11