Enabling Manual Mode

WARNING
6-2
Enabling Manual Mode
From AED Mode, press
the softkeys above the display to enter the password.
Figure 6-1: Manual Password Display
1
2
Enter Password
EXIT
PASSWORD
To return to AED Mode, without entering a password, press
Once Manual Mode is enabled, the green LED next to
In its default configuration, the HeartStart XLT powers on into AED Mode.
If you prefer, you can configure it to power on into Manual Mode. You have
the option to require users to enter a password in order to enable Manual
Mode from AED Mode. (See "Configuring the HeartStart XLT" on page 10-7.
Use of a password for Manual Mode is recommended as a safety measure to deter
untrained personnel from entering Manual Mode.
Manual
to enable Manual Mode. If prompted, use
3
4
Shocks: 3
00:00:49
is lit.
Manual
Defibrillating in Manual Mode
.