Pausing for CPR

NOTE
3-12
Pausing for CPR
At the completion of a shock series or when no shock is advised, the Heart-
Start XLT prompts you to
you to check the pulse, then prompts you as follows:
Figure 3-10 Press Pause Display
If Needed
Press PAUSE
And Begin CPR
ANALYZE
PAUSE
If CPR is needed, press
elapsed time and the total duration of the Pause state, in seconds. The Pause
Timer is configurable to meet your local CPR protocol needs. Rhythm, SpO
and heart rate monitoring alarms are suspended for the duration of the pause.
* This section describes how the Pause state functions using the default con-
figuration. If your HeartStart XLT is configured to support the European
Resuscitation Council Guidelines for Resuscitation, refer to the "ERC Proto-
col" section on page 3-16 for details.
. It allows eight seconds for
Shocks: 1
00:00:50
. While paused, the Pause Timer indicates the
Defibrillating in AED Mode
2