Aed Mode - Philips HEARTSTART XLT M3500B Instructions For Use Manual

Defibrillator/monitor
Table of Contents

Advertisement

Specifications
Defibrillator (cont.)

AED Mode

AED Energy Protocol: Fixed Energy (150 Joules).
AED Shock Series: 2, 3, or 4.
Shock Series Timer: off, 30, 60, 90, 120, 150, 180, or 210 seconds.
Text and Voice Prompts: Extensive text audible messages guide user
through protocol.
Protocol: Follows pre-configured settings. Can be configured to meet
American Heart Association and European Resuscitation Council guide-
lines.
AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock,
Lead Select, SpO
, HR Alarms, Display Contrast, ECG Gain, Volume,
2
Print Strip, Print Summary, Annotated Mark Event, Manual Mode Entry.
Indicators: LCD Display for ECG waveform and text prompts, Audio
Alerts, Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone,
13
Printer.
Armed Indicators: Charge Done Tone, Available Energy indicated on
display, Voice Message.
Patient Analysis: Per protocol, evaluates patient ECG and signal quality
to determine if a shock is appropriate and evaluates connection impedance
for proper defibrillation pad contact.
Shockable Rhythms: Ventricular fibrillation with amplitude > 100 uV
and wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity: Meets AAMI guidelines.
M3500B HeartStart XLT Defibrillator/Monitor