Philips HEARTSTART XLT M3500B Instructions For Use Manual page 163

Defibrillator/monitor
Table of Contents

Advertisement

Table 13-4 Clinical Summary - Cardioversion
Cumulative
Cardioversion Efficacy
Single shock only
<2 shocks
<3 shocks
<4 shocks
Skin "burn"
None
Mild
Moderate
Severe
Number of shocks
Cumulative delivered
energy
Skin reaction definitions: (Evaluated 24 - 48 hours after procedure)
Mild - erythema, no tenderness
Moderate - erythema, tenderness
Severe - blistering or necrosis, tenderness
M3500B HeartStart XLT Defibrillator/Monitor
The 150J SMART Biphasic waveform successfully converted far more
patients with an initial 100J shock (60% compared with 22% for the
monophasic waveform), and successfully converted patients at least as well
with a maximum energy of 200J as the monophasic did with its maximum
energy of 360J (91% compared to 85% for the monophasic waveform). Over-
all, the biphasic waveform required fewer shocks (1.7, compared to 2.8 for the
monophasic waveform) and lower delivered energy (217J, compared to 548J
for the monophasic waveform). Outcomes are summarized in , Table 13-4.
Biphasic Patients
Number (Percent)
58/96 (60%)
74/96 (77%)
86/96 (90%)
87/96 (91%)
25/90 (28%)
50/90 (56%)
15/90 (17%)
0/90 (0%)
1.7 + 1.0
217 + 176J
Clinical Performance Summary - Cardioversion
Monophasic Patients
Number (Percent)
24/107 (22%)
47/107 (44%)
57/107 (53%)
91/107 (85%)
15/105 (14%)
47/105 (45%)
41/105 (39%)
2/105 (2%)
2.8 + 1.2
548 + 331J
P Value
<0.0001
13
<0.0001
<0.0001
.29
0.0001
<0.0001
<0.0001
13-17