The report also lists additional checks that you should do. Perform each of
these checks and record the results. The guidelines for completing the checks
are as follows:
Defibrillator Inspection - make sure the HeartStart XLT is clean, clear of
objects on top and has no visible signs of damage.
Cables/Connectors - make sure there are no cracks, broken wires, or other
visible signs of damage. Make sure the connectors engage securely.
Supplies - make sure the carrying case has:
two sets of multifunction defib electrode pads in sealed packages,
within the expiration date
an adequate supply of monitoring electrodes, within the expiration
date
alcohol wipes
hand towel
scissors
a razor
an extra roll of printer paper
a spare charged battery
a Data Card
SpO
sensors (if monitoring SpO
2
Operational Checks
)
2
11
11-5