12.6 Electromagnetic compliance (EMC) - Manufacturer´s declaration
Test
Mains PortConducted Emissions
Radiated Emissions
Mains Harmonics
Voltage Fluctuations
Electrostratic Discharge
Radiated Field Immunity
EFT/Bursts
Surge Immunity
Conducted RF Immunity
Magnetic Field Immunity
Dips & Interruptions
1488727-M
Basic Standard
EN 55011
EN 55011
EN 61000-3-2 +A1 + A2
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Compliance Level / Frequency Range
Group 1 Class B
Group 1 Class B
Class A
dc%/dmax/d(t)/Pst
±2/4/6kV Contact, ±2/4/8kV Air
3V/m, 80% 1kHz AM, 80MHz - 2.5 GHz
±2kV L/N/E/LNE
±0.5/1kV L-N, ±0.5/1/2kV L-E/N-E
3V, 80% 1kHz AM, 0.15MHz - 80MHz
3A/m, 50Hz-80Hz
-95%/-60%/-30%